Contact an Account ManagerHPLC assay of hydromorphone hydrochloride injection
Summary
Type: HPLC
Application #: 101240
Catalog#: 720120.46 - MN HPLC Column, Nucleosil C18, 100A, 5um, 4.6x150mm
Detection: UV 280nm
Conditions: eluent: mixture of 20ml glacial acetic acid,
10g sodium lauryl sulphate and
800ml acetonitrile,
filled up with water to 2000ml,
filtered through 0.4μm membrane filter;
flow rate 1.5 ml/min,
injection volume 100μl
Authors: Menon S K
Source: J Liquid Chromatogr 12 (1989) 657 - 671
Details
Legend: Chromatograms of
A. a system suitability preparation,
B. typical hydromorphone sample solution and
C. a light degraded sample.